Loss-on-Drying Method Validation for a Moisture Analyzer
By Lim Xu, Export Manager
Loss-on-drying method validation for a moisture analyzer means checking the rapid method against a reference oven method on your own materials.
Loss-on-drying method validation for a moisture analyzer is the work of showing that a rapid halogen method agrees with a reference oven method on your materials, under a written program. The analyzer is not validated in the abstract. The method, the material and the settings have to be checked together.
What you are validating
A halogen moisture analyzer dries a weighed sample with a halogen lamp and reports moisture from the loss in mass. The same idea sits behind many oven loss-on-drying methods, which is why laboratories compare the two.
Validation here is not a substitute for the reference method. It is the evidence that the rapid method, at a stated temperature and end-point rule, can be used for routine work on a named material. Limits that belong on a specification, and how a finished-product limit is written, are covered on a separate page about the loss-on-drying limit. This page stays with the method check itself.
Write the reference method name, the analyzer program and the material in the same protocol. If any of those change, the previous check does not automatically still hold.
Building a comparison against the oven reference
Split a homogeneous sample and run the reference oven method and the analyzer on portions prepared the same way. Keep particle size, sample mass and delay before the test as consistent as you can.
Use more than one lot or more than one moisture level if the method will be used across a range. A single passing pair of results is a start, not a validation.
The drying oven vs halogen moisture analyzer page explains the two approaches. Use it for the principle, then write your own comparison data. We will not invent an acceptance difference or a required number of replicates here. Please confirm with us if you want help designing a check for a named material.
The KHT-HMA-01 halogen moisture analyzer lists GB/T 462 and ISO 287, which are paper moisture methods. It does not claim to comply with every oven method you may use as a reference. The instrument dries between 40 and 160 °C in 1 °C steps, weighs to 0.001 g, and accepts a maximum sample of 110 g. Those are instrument limits, not a validated program.
Want to see how a material like yours behaves on a halogen analyzer? Tell us what you need to test. We test a similar material sample and send you the video, so you do not need to ship a sample. If you prefer to send your own, we accept that too.
What to hold constant in the analyzer program
Pick one temperature, one end-point rule and one sample-mass range, and keep them for that material. Stop by a stable-mass rule, not by a clock time alone, unless the written method says otherwise.
A temperature copied from an oven method is not automatically the right analyzer setting. Oven methods and halogen methods heat the sample differently. For how to think about analyzer temperature, see halogen moisture analyzer temperature settings. If you need a starting temperature for a named material, please confirm with us.
Record the pan loading, whether the sample was spread evenly, and any pre-drying or grinding step. Those details often explain a mismatch better than the instrument model does.
Precision, range and when to repeat the check
Precision is the spread of repeat analyzer results on the same material and program. Range is the moisture span over which you have actually compared the two methods. Neither figure should be copied from another laboratory's file.
Repeat the comparison when the material, the grind, the temperature or the reference method changes, and after any maintenance that could affect heating or weighing. The interval for a periodic check should follow your laboratory quality procedure.
KHT Instruments has made packaging and lab test equipment since 2013, with a 1,000 m² factory in Jinan, China. Datasheets for KHT-HMA-01 are available on request from KHT.
Frequently asked questions
Does a factory calibration certificate validate my loss-on-drying method?
No. A calibration check supports the instrument. Method validation shows that your program agrees with the reference method on your materials.
Can I use a published acceptance limit as my validation criterion?
Only if that limit is in the method or specification you follow. We do not publish a general pass-or-fail difference between oven and analyzer results.
Is the loss-on-drying formula the same as method validation?
No. The loss-on-drying formula is how moisture is calculated from the mass change. Validation is the comparison work around that calculation.