Moisture Analyzer for Plastic Resin Before Molding

By Lim Xu, Export Manager

A moisture analyzer for plastic resin checks pellets before molding. See which resins absorb water, how ASTM D6980 works and when Karl Fischer fits.

A moisture analyzer for plastic resin checks that pellets are dry enough to process before they reach the press. ASTM D6980 describes this loss in weight method for most plastics. ASTM D6869 gives a Karl Fischer method for cases where water alone has to be measured.

Why resin moisture matters before molding

Moisture changes how some resins process. ASTM D6980 states that defects occur when these materials are processed outside the recommended moisture range. It adds that moisture greatly affects the physical properties of some plastics.

ASTM D6869 names the typical faults. High moisture causes surface imperfections such as splay or bubbling, or degradation by hydrolysis. Low moisture combined with high temperature can cause polymerization.

Hydrolysis is the harder fault to catch. Plastics Technology explains that in PET, PBT, PC, nylon and TPU, moisture at processing temperature breaks long polymer chains into shorter ones. The molded part may look fine and pass a room temperature test, then fail early when it runs hot or cold.

Resin can also pick up water before it reaches your dryer. The same source notes that resin leaves its maker dry but absorbs moisture inside the packaging, so it is already wet when the bag or container is opened.

Which resins are moisture sensitive

Resins fall into two groups. Hygroscopic resins absorb water into the pellet, while non-hygroscopic resins hold it only on the surface. For the moisture limit of a specific grade, use the technical data from your material supplier.

Hygroscopic resins

Common hygroscopic resins include nylon, ABS, acrylic, PET, PBT, polycarbonate and polyurethane. Nylon is one of the most sensitive to moisture regain. Because the water sits inside the pellet, these resins are dried before processing.

Dried pellets can pick water up again. Plastics Technology notes that dried pellets left in a hopper open to room air attract humidity again. Test pellets close to the time they go into the machine, not hours earlier.

Non-hygroscopic resins

Polypropylene, polyethylene, polystyrene and PVC are typical non-hygroscopic resins. They collect moisture only on the pellet surface.

Surface moisture still matters. Cold pellets brought from an unheated warehouse into a warm, humid plant can collect condensation when the container is opened. Filled, reinforced or carbon black compounds also hold some surface moisture, and adding a component to polyethylene or polypropylene can make the compound hygroscopic.

Not sure how your resin will behave on a halogen analyzer? Tell us your test requirements. We source similar samples, run the test and send you the video. No need to ship samples.

Can a halogen analyzer measure low resin moisture?

It can, provided the analyzer resolves the moisture level your resin specification sets. ASTM D6980 covers moisture down to 50 mg/kg (50 ppm) in most plastics.

Our KHT-HMA-01 halogen moisture analyzer weighs to 0.001 g and shows moisture in 0.01% steps. Compare those values with the limit on your resin data sheet before you use it for this job.

A loss in weight result counts every volatile, not water alone. D6980 warns that volatiles other than moisture can be given off and bias the result high. It also asks you to avoid temperatures that would release volatiles unnecessarily.

Our comparison of moisture analyzer vs Karl Fischer explains what each method actually measures. That difference decides which result your specification is asking for.

Sampling pellets and setting the drying temperature

D6980 uses a small specimen, so it asks you to make sure the specimen represents the sample. Draw pellets from a well mixed sample rather than from the top of an open bag.

Keep samples sealed until the test. D6980 asks for hygroscopic plastics to be stored in airtight containers of glass or another suitable material. Pellets that were dried before testing must cool to room temperature in a sealed container first.

Set the drying temperature by experiment. D6980 links several programs that differ only in temperature, each one 5 °C higher than the last, then plots the result against temperature. Choose the test temperature in the range where the result is very low and constant.

Above that range, the result starts to rise again. D6980 puts this down to water generated by decomposition or solid phase polymerization of the sample. Do not raise the temperature to a point where harmful fumes could form.

For resins listed in its annex tables, D6980 gives suggested test temperatures. Our guide to halogen moisture analyzer temperature settings walks through the same temperature study step by step.

When Karl Fischer or ASTM D6869 is needed

Use Karl Fischer when the specification asks for water itself, or when other volatiles leave the resin at the test temperature. ASTM D6869 determines moisture in plastics through the reaction of iodine with water. It covers a range of 0.005 to 100%.

Karl Fischer has its own limits. D6869 warns that plastics with residual monomers or plasticizers can release components that interfere with the iodine and water reaction.

Karl Fischer also helps set up the halogen test. D6980 notes that a Karl Fischer apparatus is useful when you work out the optimal test conditions for a material.

Frequently asked questions

Can a loss in weight test be used for incoming resin acceptance?

Yes. ASTM D6980 says its loss in weight method is intended for control, acceptance and assessment testing. Agree the method, drying temperature and moisture limit with your resin supplier before you use it to accept or reject a lot.

Is there an ISO equivalent to ASTM D6980?

No. D6980 states that there is no known ISO equivalent. For Karl Fischer work, ASTM D6869 is equivalent to Method B of ISO 15512.

Get a test video for your samples

Tell us your test requirements, such as the material, the size and the test you need. We source similar samples here, run the test on the instrument you are looking at and send you the video. No need to ship samples.

Questions about a model or a test? Email Amy Zhao, Export Manager, at [email protected].

Related articles

Get Your Free Quote

Looking for the right analyzer? Talk to our experts and find the best solution for your application.

View our privacy policy.

© 2026 Halogen Moisture Analyzer | All rights reserved.