Moisture Analyzer SOP Template for Halogen Instruments

By Lim Xu, Export Manager

A moisture analyzer SOP makes every operator run a halogen analyzer the same way. Use this template for scope, checks, calibration, test steps and records.

A moisture analyzer SOP is the written procedure that makes every operator run a halogen moisture analyzer the same way. It fixes the method settings, the checks before use, the test steps and how results are recorded and reviewed.

What a moisture analyzer SOP must cover

The SOP has to control the factors that move a loss on drying result. A 2014 NPL review lists sample preparation and handling, sample size, temperature, humidity and drying time among them. It names preparation and handling as possibly the largest source of uncertainty.

The method settings belong in the SOP as well. ASTM D6980 explains that the measuring conditions are developed by adjusting the analyzer's parameter settings. A result only repeats when those settings are written down and used every time.

The table below lists the key sections of a template and what each one should fix.

SOP sectionWhat to writeWhy it matters
ScopeMaterials, sample forms and the reference methodSettings are developed for each material
Checks and calibrationBalance and heater calibration, verification sample, acceptance rangeBalance and heater both affect the result
Sample handlingStorage, sampling, minimum sample massHandling can be the largest uncertainty source
Method settingsDrying temperature, program and end point for each materialThe analyzer stops when weight loss falls below a set rate
RecordsSample identity, date, specimen size, each result and the averageA reviewer can trace and repeat the test

Scope, responsibilities and safety

Start the scope with the materials and sample forms the SOP covers. ASTM D6980, for example, applies to most plastics and accepts specimens as powders, pellets or ground material.

If a product follows a pharmacopoeia, name the method the analyzer result is checked against, such as USP <731> Loss on Drying. Keep that method as the reference until the halogen result has been shown to agree with it on the same material.

Name who runs the tests, who reviews the results and who looks after calibration. Where staffing allows, the person who approves a result should not be the person who produced it.

Write the safety points into the steps. D6980 warns that specimens are hot after the test and that the drying temperature must not reach a point where harmful fumes can form. It leaves the choice of safety practices to the user.

Want to see your sample run before you write the method? Tell us your test requirements. We source similar samples, run the test and send you the video. No need to ship samples.

Daily checks and calibration

Check the analyzer before the first test of the day. It should be level, switched on for the warm-up time in its manual and placed away from draughts and vibration. These are the same preconditions the EURAMET weighing guide sets before a balance is calibrated.

Calibrate both the balance and the heater. D6980 requires both to be calibrated against a traceable standard, then verified with sodium tartrate dihydrate. That salt has a known crystal water content of 15.66%, and the acceptable result range is 15.61 to 15.71%.

If the verification fails, D6980 has you repeat it. If it fails again, run a temperature calibration and then a balance calibration, and retest. Testing stops until the cause is found and corrected.

Set a minimum sample mass as well. The EURAMET guide explains that a user can define a minimum weight, the smallest net sample mass that still meets the required weighing accuracy. Written into the SOP, it stops operators from weighing samples that are too small.

Our KHT-HMA-01 halogen moisture analyzer takes samples up to 110 g and weighs to 0.001 g. Record those values in the SOP, because the readability sets the smallest mass change the analyzer can show.

Measurement procedure

The steps below follow the order of the procedure in ASTM D6980. Use the settings validated for each material.

  1. Prepare the analyzer as its manual describes and load the program for the material.
  2. Let the heating chamber preheat and settle at its idle temperature.
  3. Start the program and follow the prompts for placing the sample on the pan.
  4. Spread the specimen in an even layer over the pan.
  5. Let the test run until the weight loss rate falls below the end point setting.
  6. Record the result in percent moisture.
  7. Let the analyzer cool if needed and remove the hot pan with care.
  8. Place a clean pan and let it settle before the next test.

For a new material, find the drying temperature before you write it into the SOP. Our guide to halogen moisture analyzer temperature settings explains the temperature study D6980 uses.

Handle samples quickly and keep them closed until weighing. The NPL review notes that laboratory humidity and temperature during preparation can make a sample gain or lose moisture.

Recording and reviewing results

Record enough to repeat the test. D6980 asks for the sample identity, including material type, source, manufacturer's code, form and previous history. It also asks for the test date, each specimen size, each specimen result and the average when several tests are run.

Add the program, the drying temperature and the analyzer identity to each record. Without them, a reviewer cannot tell whether two results were measured the same way.

Compare results with a reference method at set intervals. Where water content is what matters, the NPL review advises comparing loss on drying results with a water specific technique such as Karl Fischer titration.

Frequently asked questions

Can one SOP cover several materials?

Yes, as long as each material has its own validated settings. D6980 notes that test conditions for materials not listed in its annex tables have to be found by experiment or with help from the analyzer manufacturer. List the settings for each material in an annex to the SOP.

Should moisture be reported on a wet or dry basis?

Usually wet basis. The NPL review notes that moisture content is normally expressed as a percentage of the wet mass unless dry basis is specified. State the basis in the SOP and on every report, because the two results differ for the same sample.

Get a test video for your samples

Tell us your test requirements, such as the material, the size and the test you need. We source similar samples here, run the test on the instrument you are looking at and send you the video. No need to ship samples.

Questions about a model or a test? Email Amy Zhao, Export Manager, at [email protected].

Related articles

Get Your Free Quote

Looking for the right analyzer? Talk to our experts and find the best solution for your application.

View our privacy policy.

© 2026 Halogen Moisture Analyzer | All rights reserved.