Halogen Moisture Analyzer Calibration and Verification

By Lim Xu, Export Manager

Halogen moisture analyzer calibration covers the balance and the heater. See how each is checked, then how to verify the instrument and your method.

Halogen moisture analyzer calibration covers two parts of the instrument, the balance and the heater, and each is checked against a traceable standard. You then verify the whole instrument with a material of known water content. Method verification is a separate step that shows your drying program gives the right answer for your own sample.

What does calibrating a halogen moisture analyzer involve?

ASTM D6980 says that to maintain the integrity of test results, the balance and the heater shall both be calibrated using a traceable standard. Both matter because the moisture result is the mass lost divided by the starting mass. The balance supplies both masses, and the heater decides how much of the sample leaves.

The table below compares the four checks this article covers. They answer different questions, so passing one does not cover the others.

CheckWhat it confirmsReference used
Balance calibrationThe weighing system reads mass correctly across its rangeStandard weights
Temperature calibrationThe heater reaches the temperature the program setsA traceable temperature standard
Instrument verificationBalance and heater together give a known resultA material of known water content
Method verificationYour program measures your sample correctlyA reference method on the same sample

How is the balance calibrated?

EURAMET Calibration Guide No. 18 lists three tests normally performed on a weighing instrument. They cover the repeatability of indications, the errors of indication and the effect of placing a load off centre.

  • Repeatability, with the same load placed on the pan at least five times under identical conditions.
  • Errors of indication, with at least five test loads spread fairly evenly over the normal weighing range.
  • Eccentricity, with one test load placed at the centre and then at positions around the pan.

The errors of indication test must use standard weights or suitable substitution loads. EURAMET also says the details of a calibration can be agreed between the client and the calibration laboratory, in view of how the instrument is normally used.

Weighing accuracy limits how good the moisture result can be. ISO 287 asks for a balance accurate to 0.05% of the mass being weighed, or better, for its oven method on paper and board. Our KHT-HMA-01 weighs samples up to 110 g with 0.001 g resolution, and it uses external calibration. The halogen moisture analyzer page lists the full specification.

Questions about a model or a test? Email Amy Zhao, Export Manager, at [email protected].

How is the heating temperature calibrated?

Temperature calibration checks that the heater delivers the temperature your program asks for. D6980 calls for the heater to be calibrated against a traceable standard, alongside the balance.

The reading you see comes from a sensor, not from the sample itself. D6980 notes that the optimal test temperature for a plastic can sit above its melting point, because of the distance between the resistive thermal device and the sample pan. A heater that drifts therefore moves the result even when the balance reads correctly.

Traceability is your responsibility to document. A 2014 NPL literature review found that loss on drying standards do not address how the oven temperature and the balance are calibrated. They give only tolerances for oven temperature and balance accuracy. KHT can provide ISO/IEC 17025 traceable calibration certificates.

How do you verify the whole instrument?

D6980 verifies calibration with sodium tartrate dihydrate, which has a known crystal water content of 15.66%. The acceptable result range is 15.61 to 15.71%. Other materials with a verifiable theoretical water content are also acceptable.

The standard sets out what to do with the result, in this order.

  1. Prepare the analyzer and run the verification material as a normal analysis.
  2. If the result is outside the acceptable range, run it again.
  3. If it is still outside, perform a temperature calibration first, then a balance calibration.
  4. Retest with sodium tartrate dihydrate.
  5. If the result still fails, find and correct the cause before any further testing.

Record each verification result, so a slow drift shows up before it pushes a result out of range.

How do you verify the method for your own sample?

A verified instrument can still give a wrong answer if the program does not suit the sample. D6980 finds the optimum temperature by running linked programs that each raise the temperature by 5 °C, then plotting the result against temperature.

Most of the moisture is driven off over the lower part of the curve. The result then levels off, and D6980 says to choose a test temperature in that flat range. Above it the result starts to rise again, likely because the sample decomposes or polymerises and generates water.

Loss on drying counts every volatile, not only water. The NPL review advises comparing results with a water-specific technique such as Karl Fischer titration, and D6980 notes a Karl Fischer apparatus is useful when you set test conditions. Our comparison of moisture analyzers and Karl Fischer titration explains where each one fits.

Run both methods on the same batch and compare the results, which is how NPL describes checking one moisture technique against another. Pay attention to sample handling, which the review names as possibly the greatest source of uncertainty in loss on drying.

Frequently asked questions

How often should a halogen moisture analyzer be calibrated?

There is no single interval in D6980. The standard ties calibration to verification, so a verification result outside the acceptable range triggers temperature and balance calibration. Set a routine interval in your own procedure as well, based on how often you use the analyzer.

Can I verify with something other than sodium tartrate dihydrate?

Yes. D6980 accepts other materials with a verifiable theoretical water content for validation. Use one whose water content is documented, and record which material and value you used.