Moisture Analyzer for Powder and Hygroscopic Samples
By Lim Xu, Export Manager
A moisture analyzer for powder needs careful handling, an even layer and a checked temperature. See how to test hygroscopic powders and verify results.
A moisture analyzer for powder gives a fast loss on drying result, but powders need care before and during the test. Fine particles can gain or lose water in room air, cake or crust as they dry, and need an even layer on the pan.
Why powders are tricky to measure
A powder sample can change before it is even weighed. A 2014 NPL review of loss on drying notes that laboratory humidity and temperature play a crucial role in moisture gain or loss during sample preparation and handling. It names preparation and handling as possibly the largest source of uncertainty.
Powders can also change while they dry. NPL describes caking of fine materials and, for some biological samples, a thin film forming on the surface. Both block moisture from leaving the inside of the sample and distort the drying.
The result counts every volatile. NPL notes that loss on drying is not specific to water, and that bound water can be driven off if the drying temperature is too high.
Handling hygroscopic powder before weighing
Keep the sample sealed from the moment it is taken. ASTM D6980 asks for hygroscopic materials to be stored in airtight containers of glass or another suitable material, and it lists powders among the specimen forms it accepts.
Material that was dried or heated before the test should cool in a sealed container. D6980 asks for this, and reference oven methods such as ISO 712 and ISO 5537 cool their samples in a desiccator before weighing.
Mix the bulk sample before you take the test portion. D6980 notes that a small specimen has to be checked carefully to make sure it represents the sample.
Work quickly at the pan. Have the analyzer ready, then weigh and spread the powder in one step, so it spends as little time as possible in room air.
Not sure how your powder will dry? Tell us your test requirements. We source similar samples, run the test and send you the video. No need to ship samples.
Spreading the sample and choosing sample mass
Spread the powder in a thin, even layer across the pan. D6980 describes the specimen as spread onto the sample pan, and an even layer avoids thick spots where caking and crusting start.
Reference oven methods show the sample masses that standard methods use for common powders. The table lists three of them, taken from the NPL survey of moisture standards.
| Powder | Reference oven method | Oven sample mass | Oven drying conditions |
|---|---|---|---|
| Starch | ISO 1666 | 5 ± 0.25 g | 130 to 133 °C for 1.5 h |
| Dried milk | ISO 5537 | 5 ± 0.3 g | 87 ± 1 °C for 5 h in a ventilated oven with dry air |
| Cereals and cereal products | ISO 712 | 5 ± 1 g | 130 to 133 °C for 120 ± 5 min (90 min for flour) |
These are oven values for the reference method, not halogen settings. Use them as the method you check against, then set the halogen program by comparison on the same material.
Choose a sample mass large enough for the mass lost to be weighed reliably. The EURAMET weighing guide calls the smallest net sample that still meets the required accuracy the minimum weight. Our loss on drying formula guide shows how the result comes from the masses before and after drying.
Our KHT-HMA-01 halogen moisture analyzer takes samples up to 110 g on a 90 mm pan and weighs to 0.001 g.
Temperature and end point for powders
Pick the temperature for the powder, not for the instrument. NPL notes that the water left after drying depends on the drying temperature, and that higher temperatures can lead to chemical decomposition. Dried milk, for example, is dried at a much lower oven temperature than starch or flour in the table above.
Our guide to halogen moisture analyzer temperature settings explains how to find the temperature with a step study and confirm it.
The analyzer ends the test when the weight loss rate falls below the end point you set, as D6980 describes. Oven methods either dry for a fixed time or repeat drying until two weighings agree within a set limit. Fix the end point in your method so every operator stops at the same point.
If the sample darkens or smells burnt, the temperature is likely too high for that powder. Lower it and repeat the step study.
Checking results against a reference method
Check the halogen result against the reference method on the same batch before it replaces that method. NPL notes that faster indirect methods are checked by applying both methods to the same batch or sample and comparing the results.
For pharmaceutical powders that follow a monograph, the reference is the method the monograph names, such as USP <731> Loss on Drying.
Where the specification asks for water rather than total volatiles, compare with Karl Fischer titration. NPL recommends comparing loss on drying with a water specific technique whenever water content is the main concern.
Repeat the comparison when the powder grade, the supplier or the process changes. The settings were found for one material and may not carry over.
Frequently asked questions
Should powders be ground before testing?
Only if the reference method grinds them. Some reference methods in the NPL survey test ground samples, while others test the material as received. Prepare the halogen sample the same way as the reference sample, so the two results can be compared.
Does the sample pan material matter?
Yes. ASTM D6980 calls for sample pans made of aluminum or another nonreactive material. Place a clean pan for each test, as D6980 does, so residue from the last powder does not add to the next result.