Moisture Analyzer Accuracy and Repeatability Explained

By Lim Xu, Export Manager

Moisture analyzer accuracy depends on trueness and precision, not on display steps. Learn what limits a result and how to check repeatability in your lab.

Moisture analyzer accuracy is how close a result comes to the true moisture content of the sample. It is judged through trueness and precision, and precision is checked as repeatability and reproducibility. The balance, the sample preparation and the drying program all set how good the result can be.

What accuracy means for a moisture analyzer

In the international vocabulary of metrology (VIM), measurement accuracy is the closeness of agreement between a measured value and the true value. The VIM adds that accuracy is not a quantity and has no numerical value. A measurement is simply called more accurate when its error is smaller.

ISO 5725-1 splits accuracy into two parts. Trueness describes how close a method gets to the true result, while precision describes how well it repeats a result.

A moisture analyzer raises one more question, which is what the true value is. A 2014 NPL review notes that loss on drying is not specific to water. A halogen result can therefore repeat closely and still differ from a water specific method such as Karl Fischer titration.

Repeatability vs reproducibility

Precision always comes with conditions. The VIM defines three sets of conditions, compared in the table below.

Precision typeConditionsWhat it tells you
RepeatabilitySame procedure, operators, measuring system, operating conditions and location, over a short periodThe scatter you see in routine testing
Intermediate precisionSame procedure and location over a longer period, with possible changes such as new calibrations or operatorsHow stable your lab is from day to day
ReproducibilityDifferent locations, operators and measuring systemsHow well different labs or sites agree

Precision is usually expressed as a standard deviation, a variance or a coefficient of variation under those conditions. Always state which conditions a precision value belongs to.

Want to see how a sample like yours repeats? Tell us your test requirements. We source similar samples, run the test and send you the video. No need to ship samples.

What limits the result

Three things set the limit for most moisture results. They are the balance, the way the sample is prepared and the drying program.

Balance readability

The balance sets the smallest change the analyzer can show. The VIM defines the resolution of a displaying device as the smallest difference between indications that can be meaningfully distinguished.

Readability is not accuracy. A small sample loses only a little mass, so the same weighing step becomes a larger share of the result. The EURAMET weighing guide handles this with a minimum weight, the smallest sample quantity that still achieves a specified relative weighing accuracy.

Our KHT-HMA-01 halogen moisture analyzer weighs to 0.001 g and shows moisture in 0.01% steps.

If you compare versions of our fully automatic halogen moisture analyzer, check the readability listed for each version against the sample mass you plan to use.

Sample preparation

The NPL review names sample preparation and handling as possibly the largest source of uncertainty in loss on drying. ISO 5725-1 also notes that powders and other test items need due consideration of any heterogeneity before precision is assessed.

Drying program

Temperature, end point and time all move the result. NPL notes that the water left after drying depends on the drying temperature, while D6980 ends the test when the weight loss rate falls below a set value. D6980 also defines lift, where convection currents raise the pan and falsely show a moisture loss.

How to check repeatability in your lab

Run replicate tests under repeatability conditions. Use one sample batch, one operator, one analyzer and one program, all within a short period.

  1. Mix one batch of sample well and divide it into several portions.
  2. Test the portions one after another with the same program.
  3. Calculate the mean and the standard deviation of the results.
  4. Repeat the series on another day or with another operator.
  5. Compare the two series to see your intermediate precision.

Check trueness separately, with a material of known water content. D6980 verifies analyzer calibration with sodium tartrate dihydrate, which has a known crystal water content of 15.66%, and accepts results from 15.61 to 15.71%.

Then check the result against your reference method on the same material. Our comparison of drying oven vs halogen moisture analyzer explains why that equivalence has to be shown before one method replaces the other.

Reading a moisture analyzer specification

Read each number for what it is. Readability is the smallest weight step the balance shows, and moisture resolution is the smallest step in the displayed result. Neither one is the accuracy of a moisture result.

Ask how any repeatability value was measured, including the sample, the sample mass and the number of tests. ISO 5725-1 assumes the identical item is measured in exactly the same way, so a value from a stable test material may be tighter than what your product gives.

Treat a single accuracy number with care. Under the VIM, accuracy has no numerical value, so ask for trueness and precision data and the conditions behind them.

Frequently asked questions

What repeatability can I expect from a moisture analyzer?

It depends on the material and its moisture level. In the ASTM D6980 interlaboratory study of plastics, the repeatability standard deviation ranged from 0.0024 to 0.0241 in percent moisture across the materials listed in the public preview of the standard. Measure your own value on your own product.

Does room humidity affect moisture analyzer results?

Yes. The NPL review notes that during sample preparation and handling, lab humidity and temperature play a crucial role in moisture gain or loss. Keep samples closed until weighing, and place the analyzer away from draughts, which the EURAMET guide lists among weighing influences.

Get a test video for your samples

Tell us your test requirements, such as the material, the size and the test you need. We source similar samples here, run the test on the instrument you are looking at and send you the video. No need to ship samples.

Questions about a model or a test? Email Amy Zhao, Export Manager, at [email protected].

Related articles

Get Your Free Quote

Looking for the right analyzer? Talk to our experts and find the best solution for your application.

View our privacy policy.

© 2026 Halogen Moisture Analyzer | All rights reserved.