Loss on Drying Limit for Tablets, Granules and APIs

By Lim Xu, Export Manager

Where a loss on drying limit comes from, how a QC lab applies it to tablets, granules and APIs, and when a halogen method needs comparison.

A loss on drying limit is an acceptance criterion for the mass a material may lose under a specified drying procedure. It is not a universal moisture limit for all tablets, granules or active pharmaceutical ingredients (APIs). To use a result for release, first identify the applicable monograph or approved specification, then run the method that specification requires.

Where does an LOD limit come from?

The limit and the method belong together. A pharmacopoeial monograph may give an acceptance criterion and point to a general chapter, while an approved in-house specification can define a justified criterion for a particular product. The same percentage can have a different meaning if drying temperature, time, sample preparation or end point changes.

USP <731> and Ph. Eur. 2.2.32 are distinct pharmacopoeial loss-on-drying chapters; use the chapter named by the applicable monograph.

USP <731> defines loss on drying as volatile matter driven off under its specified conditions. That includes more than water when other volatile material is present.

ICH Q6A discusses when a loss on drying procedure can be adequate and when a water-specific procedure is preferable. These sources do not establish one numerical limit for every pharmaceutical material.

Monograph limits and in-house specifications

Start with the material identity, dosage form and market. Check the current, applicable monograph and the approved specification before adopting a limit from another product. If the specification was developed in-house, keep the data and reasoning that justify the criterion and the chosen procedure.

A useful QC worksheet records the specification source, version, sample identity, drying method, calculation and decision rule. This makes it possible to explain why a result passes or fails without relying on an undocumented “typical” number.

Questions about a model or a test? Email Amy Zhao, Export Manager, at [email protected].

Checkpoints for tablets, granules and APIs

Granules can change during drying, storage and transfer before compression. Tablets may contain ingredients with different volatile behavior. An API may be hygroscopic or may contain bound water. These differences affect whether loss on drying is the right measurement and where the sample is taken.

Plan the sampling point around the manufacturing decision: incoming material acceptance, an in-process check or finished product release. Record how the sample was protected before weighing. A method intended for one of these checkpoints should not be assumed suitable for another without a documented comparison.

Can a halogen analyzer replace the specified method?

Do not treat a shorter halogen run as interchangeable with a monograph procedure by default. A halogen instrument heats and weighs in one unit; the specified procedure may use another heating arrangement, sample mass or drying end point. Compare results on the actual material across the range relevant to the limit, including samples close to the acceptance boundary.

If you are considering our halogen moisture analyzer for an LOD workflow, send KHT the material, applicable method and required report fields. We will confirm whether the instrument configuration can support the proposed check. Until that is confirmed, this article makes no claim that the model replaces a pharmacopoeial method.

Documenting method comparison

If your team needs the calculation itself, see our loss on drying formula guide.

Write down the reference procedure and proposed halogen settings before testing. Prepare matched portions, record sample handling and run repeat measurements. Review any systematic difference and the spread of results, particularly near the specification limit. Keep a rule for investigating disagreement rather than applying an unexplained correction factor.

ICH Q6A places the test procedure and acceptance criterion within the product specification. That principle matters here: a faster measurement is useful only when the lab can show that its result supports the same decision for this material.

Further reading and methods

Frequently asked questions

Is an LOD result the same as water content?

Not always. USP <731> describes loss of volatile matter under the specified conditions. A water-specific method may be needed where other volatile components or water of hydration affect the decision.

Can I use the LOD limit from another tablet?

No. Use the applicable monograph or approved product specification and its defined method. A value from another formulation is not an acceptance criterion for yours.

Discuss the method with KHT

We sell direct from our factory, so you work directly with our factory team.

Questions about a model or a test? Email Amy Zhao, Export Manager, at [email protected]. Get a Quote with your material, specification and proposed test procedure.

Related articles

Get Your Free Quote

Looking for the right analyzer? Talk to our experts and find the best solution for your application.

View our privacy policy.

© 2026 Halogen Moisture Analyzer | All rights reserved.