How to Use a Halogen Moisture Analyzer for a First Run

By Lim Xu, Export Manager

How to use a halogen moisture analyzer on a first run: load the pan, choose a trial program, and read where the result levels off before you write an SOP.

How to use a halogen moisture analyzer comes down to a few choices on the first runs: how the sample goes on the pan, which trial program you start with, and how you read the result. This guide covers that first-run stage. Once your settings repeat, a written SOP takes over the daily routine.

What the analyzer measures and what you need first

A halogen moisture analyzer heats a weighed sample with a halogen lamp and reports moisture from the loss in mass. It is a rapid lab method. For many materials the oven method remains the reference, as explained in drying oven vs halogen moisture analyzer.

Before the first run, you need a clean sample pan, a representative sample of one material, a trial program (temperature and end-point rule), and somewhere to note each result. The KHT-HMA-01 halogen moisture analyzer takes a maximum sample of 110 g, weighs to 0.001 g, and dries between 40 and 160 °C in 1 °C steps. Moisture is reported from 0.00 to 100.00 % with a resolution of 0.01 %.

Those figures are instrument limits, not a method for a named product. If you are moving an oven method onto the analyzer, compare the two on the same lot before you rely on the rapid result.

Load the sample on the pan

Keep the sample closed until you are ready to weigh it. Hygroscopic powders, resins and ground foods can gain or lose water from room air while they wait.

Tare a clean, dry pan on the analyzer, then spread the sample in a thin, even layer. A heap in the middle dries slowly, and a thick cake can scorch on top while the inside is still wet.

On the first runs, use a similar sample mass each time. That way a change in the result comes from the program, not from the load. If the material has to be ground, grind it just before the run and note it, so later runs are prepared the same way.

Not sure how your sample will behave on a halogen analyzer? Tell us what you need to test. We test a similar material sample and send you the video, so you do not need to ship a sample. If you prefer to send your own, we accept that too.

Choose a trial program

For a material you have not run before, start with one drying temperature and one end-point rule, and change only one of them between runs. The analyzer sets temperature in 1 °C steps between 40 and 160 °C. A hotter program finishes sooner, but it can also drive off other volatile matter or burn the sample.

Tie the end point to a stable mass, not only to a clock. Once you change the rule, the result belongs to a different method.

For the thinking behind the temperature choice, see halogen moisture analyzer temperature settings. If you do not yet have a starting program for a named material, please confirm one with us rather than borrowing settings from another product.

Run the test and read where the result levels off

Close the chamber, start the program, and leave the lid shut until the run ends. Opening it mid-run lets in room air and disturbs the mass reading.

On the first runs, look at how the result arrives, not only at the final number. KHT-HMA-01 has a timer from 1 to 99 minutes, so you can stop trial runs on the same lot at different times and see where the moisture value stops changing.

If the value keeps creeping up and never settles, the temperature may be driving off more than water, or the sample may be breaking down. Try a lower temperature on the same lot. If two temperatures each give a stable but clearly different value, compare both with the oven result before you choose.

If one run is far from the others, check the layer, the pan and whether the lid was opened, then repeat it before you change anything else.

From first runs to a written SOP

When repeat runs on the same lot agree, note the settings that produced them: the temperature, the end-point rule, the sample mass and how the sample was prepared. KHT-HMA-01 can keep a set of programs in memory, and it has an RS232 port if you send results to a computer.

Those settings then go into your lab procedure. Our moisture analyzer SOP template covers daily checks, calibration, the full test steps and how results are recorded and reviewed, so this guide does not repeat them.

KHT Instruments has made lab instruments since 2013, with a 1,000 m² factory in Jinan, China. Datasheets for KHT-HMA-01 are available on request from KHT. The standards listed for this model are GB/T 462 and ISO 287, both paper moisture methods.

Frequently asked questions

Can I open the lid to stir a sample mid-test?

No. Opening the chamber lets in room air and changes the mass reading. If a sample needs a special preparation, do that before you start the drying program.

How many trial runs do I need before I fix a program?

Enough to show that repeat runs on the same lot agree at the settings you plan to keep. Run the oven method on the same lot as well if it is your reference.

Does a test video replace a quotation?

No. The video shows how a similar sample behaves on the instrument. A quotation still needs the model, the options and the shipping details you ask us to include.

Get a test video for your samples

Tell us what you need to test, such as the material, the size and the test method. We test a similar material sample on the instrument you are looking at and send you the video, so you do not need to ship a sample. If you prefer to send your own, we accept that too.

Questions about a model or a test? Email Amy Zhao, Export Manager, at [email protected].

Related articles

Get Your Free Quote

Looking for the right analyzer? Talk to our experts and find the best solution for your application.

View our privacy policy.

© 2026 Halogen Moisture Analyzer | All rights reserved.